Mobile:+86 15890068607
Autoclave steam sterilizers, commonly known as high-pressure steam sterilizers, are a common type of equipment that uses high-temperature and high-pressure steam to achieve sterilization. They are mainly divided into two types: portable high-pressure steam sterilizers and vertical high-pressure steam sterilizers.

Pressure steam sterilizer is widely used sterilization equipment in laboratories, hospitals, factories, and other environments. The four key parameters of the steam sterilization process are steam, pressure, temperature, and time. What to do if the temperature control of the steam sterilizer is insensitive?
If the temperature does not rise during sterilization, this is often due to insufficient pressure. If the pressure in the jacket or inner chamber is insufficient, the pressure reducing valve can be used to adjust the pressure in the jacket to the required level. If the pressure in the inner chamber is sufficient but the temperature does not rise, it indicates that there is water in the sterilizer chamber. The bypass channel to the inner chamber can be opened to drain the water for 10-20 seconds.

Steam sterilizer is a device that use pressurized saturated steam to quickly and reliably disinfect and sterilize items. They have advantages such as high sterilization efficiency and strong penetration ability, and are suitable for medical and health institutions, scientific research institutions, agriculture and other units.
Previous: No Information