Horizontal cylindrical pressure steam sterilizer
Characteristics about 500L Horizontal Cylindrical Steam Sterilizer:
1. Sterilizing course automatic sterilization controlled,easy to operate.
2. With drying function,suitable dressing drying.
3. With over-temperature,over-pressure auto-protect device.
4. The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.It can't be start on if the door doesn't close well.
5.The safety value will be open automatically when the inner pressure over 0.24MPa,and the steam be exhausting to the water tank.
6. The power will be cut-off automatically,cut-off the water and alarm when the machine lacking water.
7. The chamber of the sterilizer is made of stainless steel.
Technical data about 500L Horizontal Cylindrical Steam Sterilizer:
Technical data | WS-90YDA | WS-150YDA | WS-200YDA | WS-280YDA | WS-400YDA | WS-500YDA |
Sterilizing chamber volume |
90L φ440×700mm |
150L φ440×1000mm |
200L φ515×1000mm |
280L φ600×1000mm |
400L Φ700×1100mm |
500L Φ700×1300mm |
working pressure | 0.22MPa | |||||
working temperature |
134℃ | |||||
Adjustment of temperature | 105℃ - 134℃ | |||||
Time of sterilization | 0-60 min | |||||
Time of drying | 0-60 min | |||||
Heat average | ≤±1℃ | |||||
Power | 9KW/380V 50Hz | 9KW/380V 50Hz | 9KW/380V 50Hz | 12KW/380V 50Hz | 18KW/380V 50Hz | 18KW/380V 50Hz |
Dimension | 1100×600×1300 | 1400×600×1300 | 1400×670×1650 | 1400×770×1780 | 1430×880×1830 | 1800×900×1820 |
Transport dimension | 1250×750×1850 | 1550×750×1850 | 1560×820×1850 | 1680×920×2100 | 1800×1050×2100 | 1850×1050×2100 |
G.W/N.W | 270/200 KG | 320/240 KG | 350/260 KG | 465/365 KG | 530/520 KG | 580/470 KG |
We can also supply:
Electric or LPG heated | Electric Heated | Table top type |
|
|
|
Please click to details | Please click to details | Please click to details |
Hand round automatic | Digital display automation | Horizontal cylindrical type |
|
|
|
Please click to details | Please click to details | Please click to details |